Written by Rhiannon Anderson and Judith Buckland

Nuclear Accreditation QI Program

This month we reviewed radiation safety and patient-centered nuclear cardiology protocols. To close out our March focus on nuclear cardiology, this week we will review IAC nuclear accreditation requirements for on-going Quality Improvement (QI) and provide easy steps to creating your own program.

nuclear accreditation quality improvement

Quality Improvement

The quality improvement program was updated in 2016.  Facilities must have a quality improvement process in place to evaluate Quality Improvement (QI) measures.  The assessments are an ongoing process with monthly or quarterly review of results. The facility must have a minimum of two QI meetings per year. At least one of those meetings must be used to discuss the QI analyses and any QI-related topics. The IAC Nuclear QI program must include the following measures:

  • Test appropriateness
  • Technical quality and safety of the imaging
  • Interpretive quality review
  • Report completeness and timeliness
  • Correlation

Test Appropriateness

The facility must evaluate the appropriateness of each test. This requires checking that your studies are ordered for clinically necessary indications.

AUC should be measured using consecutive exams over a two-three week period in order to evaluate 5% of the annual volume. For smaller volume facilities, use a minimum of 30 consecutive exams. The Centers for Medicare and Medicaid (CMS) is in the process of implementing an aspect of the Protecting Access to Medicare Act (PAMA) of 2014 that will introduce a new framework for applying AUC. In 2018, PAMA will require ordering physicians to consult AUC using a CMS-approved CDSM with nuclear cardiology.  We will provide additional updates on this CMS requirement as more information is made available.

Technical Quality Review

IAC provides many different suggestions for technical quality review and these can be reviewed in their standards.  One of the measures listed is ‘dose verification of administered radioactive dose to prescribed dose listed in protocol’.   This is a relatively easy measure to complete:

  1. Identify 30 consecutive patients (you may select less in a smaller lab)
  2. Document the isotope dose that the patient received
  3. Note if this is within +/- 10% of your protocol dose

Radiation Safety Tips

  1. Check to see if there is a lower acceptable dose of the isotope you use to improve radiation safety
  2. If you dosing amount falls out of the =/- 10% range, identify the issue and then repeat the measure

Always keep patient safety front and center. Review our recent articles related to nuclear radiation safety.

Interpretive Quality Review

Your QI Plan must also include evaluating the quality and accuracy of Interpretation.  Two ways to accomplish this measure would be through peer review and/or clinical correlation. There are many ways to complete physician peer review (peer review forms, physician meetings, assigned through your PACs system), just find the one that is easiest for you.

Clinical Correlation

We always remind our clients that there is not a whole lot gained from correlating normal nuclear studies with normal cath results. Please try and identify abnormal nuclear studies to correlate with cardiac cath, cardiac MRI, CTA, etc.

Report Completeness & Timeliness

All facilities must track report completeness and timeliness.  There is no set number of studies that must be monitored.  Your facility may choose how many as long as this is an ongoing measure.  Your review should include ensuring final reports include all mandatory reporting components and that the turnaround time is compliant.

All MPI interpretations must be available within one working day of the examination.  (An interpretation may be in the form of paper, digital storage or accessible voice system.)  

Report Completeness Updates

All accredited facilities should have standardized reporting in place.  The 2016 IAC Nuclear Standards update included additions to some mandatory components of a complete MPI report and include:

  • Functional Results:
    • Quantitative left ventricular (LV) ejection fraction
    •  Overall LV function (normal, reduced [mild, moderate or severe] or hyperdynamic)
    • Regional wall motion abnormalities
    • All perfusion defects must be described, when a defect is due to attenuation/artifact this must also be identified as such
  • Final impressions should include:
    • Summary of perfusion (normal, equivocal, abnormal, ischemia, or infarction)
    • Summary of function (normal, equivocal, abnormal)
    • Should resolve any discrepancies

Summary

We hope you enjoyed our quick overview of how to implement a compliant quality improvement program into your nuclear cardiology lab. If you ever need assistance just contact us! We can provide a free phone consultation to answer any questions and address any hurdles you are facing within your lab. We are also available for hire through one of our QI services. We love helping improve diagnostic imaging through accreditation and education.

Judith Buckland, MBA, RDCS, FASE

intersocietal accreditation echo nuclear Learn more: Nuclear Cardiology Quality Improvement Services

References:

Intersocietal Accreditation Commission Nuclear Accreditation Standards and Guidelines, 2016 (latest standards at the time of this article)

accreditationALARAIAC StandardsICANLinterpretive qualityintersocietal accreditation commissionnuclearnuclear cardiologyQI maintenancequality improvementreport completenesstechnical qualitytest appropriatenessturnaround time

Our Services

LET US KNOW WHAT YOU THINK...

Your email will not be displayed.

Required *

Accreditation
services

Accreditation Help
Online Echo CMEs
Echo Accreditation Software
Free
Consultation

Search
for

Blog
Categories

Blog
Archive

Tags
Used

Show tags
Need Help? Share